Location: Start  >  Animal COMP ELISA >  Performance characteristics

Performance characteristics

Animal COMP® ELISA, 14-2004-86

Detection limit

The detection limit is <0.2 U/l.

Parallelism

Serum samples were diluted 5, 10, and 20 times using the sample diluent. The observed Obtained/Expected values for 1/5 to 1/20 were in the range 94-114%.

Precision

Each sample was analysed in tetraplicates on six different occasions (see table below).

Coefficient of variation

Pat.
Mean value
U/I
within
assay%
between
assay%
total
assay%
1
2.8
9,8
6,0
11,5
2
3.5
9,9
5,9
11,5
3
1.7
8,5
7,6
11,5
© 2010, AnaMar AB, Kungsportsavenyn 22, SE-411 36, Göteborg, Sweden
Phone: +46 31 732 4140, Fax: +46 31 732 41 49, info@anamar.com